Hazard notice: Blocked oxygen humidifier!!


1 Department of Anaesthesiology, BLDE University and Research Centre, Bijapur, Karnataka, India
2 Department of Anaesthesiology, Pravara Institute of Medical Sciences, Loni, Maharashtra, India
3 Department of Periodontology, Rural Dental College, Loni, Maharashtra, India
4 Department of Anaesthesiology and Critical Care, Mahatma Gandhi Medical College and Research Institute, Puducherry, India

Date of Web Publication11-Feb-2015

Correspondence Address:
Akshaya N Shetti
Department of Anaesthesiology and Critical Care, Pravara Institute of Medical Sciences, Loni, Maharashtra
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0259-1162.150171

Rights and Permissions

How to cite this article:
Karigar SL, Shetti AN, Mustilwar RG, Singh DR. Hazard notice: Blocked oxygen humidifier!!. Anesth Essays Res 2015;9:133-4

How to cite this URL:
Karigar SL, Shetti AN, Mustilwar RG, Singh DR. Hazard notice: Blocked oxygen humidifier!!. Anesth Essays Res [serial online] 2015 [cited 2021 Apr 17];9:133-4. Available from: https://www.aeronline.org/text.asp?2015/9/1/133/150171

Sir,

Oxygen therapy is one of the key components of treatment for patients in high dependency unit or intensive care unit. Different techniques are used for oxygen delivery, and one of the most commonly used methods is by simple mask or Venturi mask. Hence, it is always important to have functioning oxygen humidifier assembly to supply oxygen. Possibility of human errors, malfunctioning of equipment has been reported earlier. [1],[2] We report a rare scenario where nozzle of oxygen humidifier was blocked.

A patient with pneumonia was given supplemental oxygen as she was tachypneic and not maintaining adequate saturation with room air. There was no improvement in the saturation even after supplementation of humidified oxygen by mask; instead there was fall in saturation with increase in respiratory rate. Referral was given to anesthesiologist for elective intubation and mechanical ventilation. On examination, the moisture or fog content inside the mask with tachypnea made us to suspect absence of oxygen delivery to the patient as with our previous experience. [2],[3] When we checked the oxygen humidifier, we noticed proper movement of bobbin, an adequate level of distilled water, and proper bubble formation in humidifying chamber but absence of oxygen flow through the nozzle. There was absolutely no flow of oxygen even after increasing the flow to 12 L/min [Figure 1]a. This made us to suspect blockade of nozzle and on testing the nozzle was blocked [Figure 1]b. The humidifier was submitted to the biomedical department where they could dislodge the block that was present in the nozzle [Figure 1]c. After the removal of the block, the oxygen flow could be appreciated even with minimal flow [Figure 1]d. The oxygen saturation and respiration pattern of the patient improved after replacing with the proper functioning humidifier hence invasive ventilation was avoided in this patient. We couldn't identify the cause of nozzle blockade; it could have been due to improper storage of the equipment that leads to dirt accumulation. The key message here is; one should have high index of suspicion of the possibility of equipment error that could avoid unnecessary invasive ventilation and unnecessary investigations like arterial blood gas analysis. It is also important to store lifesaving medical equipment in a proper way.
Figure 1: Flow of O2 before (a) and after (b) releasing the obstruction (c) obstructed nozzle and (d) dirt material

Click here to view


 
   References Top

1.
Karigar SL, Kunakeri S, Shetti AN. An unexpected error in oxygen humidifier. Saudi J Anaesth 2014;8:567-8.  Back to cited text no. 1
[PUBMED]  Medknow Journal  
2.
Shetti AN, Dhulkhed VK, Roshansingh D, Khyadi S. Mechanical errors in oxygen humidifier. Saudi J Anaesth 2014;8:302-3.  Back to cited text no. 2
[PUBMED]  Medknow Journal  
3.
Shetti AN, Kadam AB, Bhatkhande SJ, Mustilwar RG. Hazard notice; improper connections and damaged outer tube in oxygen humidifier. Anesth Essays Res 2014;8:416-7.  Back to cited text no. 3
  Medknow Journal  


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed2174    
    Printed16    
    Emailed0    
    PDF Downloaded62    
    Comments [Add]    

Recommend this journal